Medivent Premium

It is an inhalation anaesthesia machine with continuous fresh-gas flow for patients. It is intended for use in operating theatres, induction and recovery rooms. It can be used with O2 and N2O supplied from a central supply system and/or from medical gas cylinder and hypo-oxy-guard.

Get Quotation

Features

In the ORC a slave valve pneumatically linked to two pressure balancing diaphragms of Oxygen & Nitrous Oxide guarantees delivery of minimum 25% Oxygen Concentration which can be raised anywhere in the range of 25% to 100% .
MEDISYS OFPD works as a true fail-safe system allowing flow of Nitrous Oxide in presence of Oxygen at Rotameter level. It locks the flow of Nitrous Oxide in the absence of flow of Oxygen through its Rotameter.

Specification

Rotameters Rotating Bobbin Flowmeters taper tubes, accurately calibrated in double/tripe scale to ensure accuracy and clarity in reading. Long - 230mm, Oxygen - 100,ml/min to 8 LPM, Nitrous - 200ml/min to 12 LPM
Ventilator Driving Source A quick-fit system is provided for driving ventilator
Patient Circuit A) Standard Magill's Circuit - 01, B) Heidbrink Valve - 01, C) Bag-Mount - 01, D) Antistatic Face Mask - 02 Nos. (Adult & Child), E) 2 Ltrs. Rebreathing Bag -02 nos. F) Complete Closed Circuit 01 no. table top: guarded table top. Provided with a Stainless Steel Tray
Pressure Regulators Redesigned heavy duty Pressure Regulators for efficacy and reliability - 02 nos. each for Oxygen & Nitrous Oxide